Electroanalysis · 1992 · 97 citations · 9 references
EngineeringBiochemistryToxicologyBiosensorsEnzymologySensorsBiological SensorsChemical SensorsElectrochemical BiosensorsEnvironmental BiosensingEnzymatic BiosensorsEnzyme-based BiosensorsCarbamate PesticidesEnzymatic Redox ReactionsEnzymatic SensorsEnzyme-based SensorsHydrogen Peroxide BiosensorsOrganophosphate BiosensorsOrganophosphate PesticidesPesticide Detection
Abstract An enzyme sensor for the determination of organophosphate and carbamate insecticides is described. The mechanism is based on the coimmobilization and sequential enzymatic reactions of acetylcholinesterase and choline oxidase. The inhibition of acetylcholinesterase is measured as decreased formation of hydrogen peroxide (H 2 O 2 ) as monitored by a H 2 O 2 sensor. As low as 10 nM paraoxon can be determined with this technique. The long‐term stability of the sensor is two months, when stored at 4°C in phosphate buffer, and one year in a dry state.
9
A new and rapid colorimetric determination of acetylcholinesterase activity
George L. Ellman, K. Diane Courtney, Valentino Andres et al. · Biochemical Pharmacology · 1961 · 26.9K citations
Enzymatic Determination of Organic Phosphorus Insecticides
P. A. Giang, S. A. Hall · Analytical Chemistry · 1951 · 85 citations