Design studies of a high resolution PET detector using APD arrays
98
Citations
12
References
2000
Year
EngineeringSensor ArrayPet-mriApd ArrayApd ArraysImage SensorPositron Emission TomographyDetection TechnologyInstrumentationRadiation ImagingNuclear MedicineRadiologyHealth SciencesElectrical EngineeringRadiation DetectionMedical ImagingDetector ArrayDesign StudiesDetector PhysicOptoelectronics
The authors evaluated a compact, high resolution PET detector module using avalanche photodiode (APD) arrays to replace bulky position sensitive PMTs. The newly developed APD array is a planar processed 4/spl times/4 array which has a 2/spl times/2 mm/sup 2/ pixel size with 0.4 mm gaps between pixels, about 60% quantum efficiency at 420 nm wavelength, and uniform high gain (>1000) across all channels. A 4/spl times/4 array of 2/spl times/2/spl times/10 mm/sup 3/ LSO crystals was coupled to an APD array. Different readout electronics and signal multiplexing schemes were explored. All crystals in the detector array were clearly identified in the flood source histogram, with average peak-to-valley ratios of about 12:1 using a charge sharing resistor network. The energy resolution was measured to be /spl sim/14% at 511 keV in the detector array. The measured timing resolution was 2.6 ns in coincidence with a LSO/PMT detector. By optimizing the readout electronics currently being used, it is likely that detector performance can be further improved. The authors have also determined depth-of-interaction (DOI) by reading out two APD arrays connected to the ends of a 2/spl times/2/spl times/22 mm/sup 3/ LSO crystal. Preliminary measurements show good DOI measurement capability with DOI positioning uncertainty between 4 and 6.5 mm.
| Year | Citations | |
|---|---|---|
Page 1
Page 1