Determination of AC-DC difference in the 0.1-100 MHz frequency range

Joseph R. Kinard, Tao Cai

IEEE Transactions on Instrumentation and Measurement · 1989 · 48 citations · 4 references

DOIFull text

Open access

Concepts

Abstract

Thermal voltage converter structures have been modeled theoretically and studied experimentally to determine their AC-DC differences in the 0.1-100-MHz frequency range. Estimated uncertainties, corresponding to the one standard deviation confidence level, for these AC-DC differences vary from 20 p.p.m. at 1 MHz to 2000 p.p.m. at 100 MHz.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

4