Mechanical performance comparison between RotaFlow and CentriMag centrifugal blood pumps in an adult ECLS model
43
Citations
14
References
2010
Year
The results obtained in this experiment demonstrate that the RotaFlow has a higher shut-off pressure (less retrograde flow) and maximal blood flow rate than the CentriMag blood pump. Findings support the conclusion that the RotaFlow disposable pump head has a better mechanical performance than the CentriMag. In addition, the RotaFlow disposable pump is 20-30 times less expensive than the CentriMag.