Concepedia

Publication | Open Access

Comparison of a sidestream capnograph and a mainstream capnograph in mechanically ventilated dogs

42

Citations

10

References

2002

Year

Abstract

RelevancehAlthough the mainstream cas slightly more accurate, both methods of ETCO2 measurement correlated well with PaCO2 and reflected changes in the ventilatory status. However, ETCO2 values > 45 mm Hg may inaccurately reflect the severity of hypoventilation as PaCO2 may be underestimated during conditions of hypercapnia (PaCO2 > 60 mm Hg).

References

YearCitations