Concepedia

Publication | Open Access

<i>In vitro</i> and <i>in vivo</i> repeatability of abdominal diffusion-weighted MRI

86

Citations

24

References

2012

Year

Abstract

Good in vitro repeatability of ADC measurements provided a sound basis for in vivo measurement. In vivo variability is higher and when considering single measurements in the abdomen as a whole, only changes in ADC value greater than 23.1% would be statistically significant using a two-dimensional ROI. This value is substantially lower (7.9%) if large three-dimensional VOIs are considered.