On-body calibration and processing for a combination of two radio-frequency personal exposimeters

Arno Thielens, Sam Agneessens, Leen Verloock, Emmeric Tanghe, Hendrik Rogier, Luc Martens, Wout Joseph

Radiation Protection Dosimetry · 2014 · 38 citations · 22 references

DOIFull text

Open access

Concepts

TL;DR

The study recommends using multiple PEMs and reporting multivariate statistics (mean vector and covariance matrix) alongside univariate summaries in future PEM research. The authors calibrated two hip‑mounted PEMs in an anechoic chamber to characterize response and crosstalk for realistic polarized fields, then performed simultaneous measurements in Ghent, Belgium. Averaging two PEMs reduced the 50 % confidence interval by 2.6 dB, but significant crosstalk (up to 1.2) can obscure measurements; measured exposures ranged from 0.052 mW/m² for GSM downlink to 0.061 µW/m² for UMTS uplink.

Abstract

Two radio-frequency personal exposimeters (PEMs) worn on both hips are calibrated on a subject in an anechoic chamber. The PEMs' response and crosstalk are determined for realistically polarised incident electric fields using this calibration. The 50 % confidence interval of the PEMs' response is reduced (2.6 dB on average) when averaged over both PEMs. A significant crosstalk (up to a ratio of 1.2) is measured, indicating that PEM measurements can be obfuscated by crosstalk. Simultaneous measurements with two PEMs are carried out in Ghent, Belgium. The highest exposure is measured for Global System for Mobile Communication downlink (0.052 mW m(-2) on average), while the lowest exposure is found for Universal Mobile Telecommunications System uplink (0.061 μW m(-2) on average). The authors recommend the use of a combination of multiple PEMs and, considering the multivariate data, to provide the mean vector and the covariance matrix next to the commonly listed univariate summary statistics, in future PEM studies.

References

22