Concepedia

Publication | Open Access

Routine evaluation of left ventricular diastolic function by cardiovascular magnetic resonance: A practical approach

162

Citations

19

References

2008

Year

Abstract

We have shown that there is homology between CMR and TTE for the assessment of diastolic inflow over a wide range of conditions, including normal, impaired relaxation and restrictive. There is excellent agreement of quantitative velocity measurements between CMR and TTE. Diastolic blood flow assessment by CMR can be performed in a single scan, with times ranging from 20 sec to 3 min, and we show that there is good indication for applying CMR to assess diastolic conditions, either as an adjunctive test when evaluating systolic function, or even as a primary test when TTE data cannot be obtained.

References

YearCitations