Publication | Open Access
Evaluation of simple rapid HIV assays and development of national rapid HIV test algorithms in Dar es Salaam, Tanzania
103
Citations
11
References
2009
Year
An alternative confirmatory HIV testing strategy based on initial testing on either SD Bioline or Determine assays followed by testing of reactive samples on the Determine or SD Bioline gave 100% sensitivity (95% CI; 99.1-100) and 100% specificity (95% CI; 96-99.1) with Uni-Gold as tiebreaker for discordant results.