A new technique for decomposition of selected ions in molecule ion reactor coupled with ortho-time-of-flight mass spectrometry
75
Citations
0
References
1997
Year
EngineeringBiological Mass SpectrometryIon Mobility SpectrometryChemistryIon ProcessChemical EngineeringMolecule Ion ReactorAnalytical ChemistryNew TechniqueIon EmissionBiophysicsChromatographyIon ExchangePhysical ChemistryOrtho-time-of-flight Mass SpectrometryIon MobilitySelective IonMass SpectrometryMedicineChemical KineticsMolecular FragmentationIon Structure
A molecule ion reactor (MIR), i.e. a gas filled radio-frequency only quadrupole with a longitudinal electrical field (RFQLEF), is used as an atmospheric pressure ionization interface for an orthogonal time-of-flight mass spectrometer (O-TOFMS). A new phenomenon of selective ion 'heating' in a MIR near Mathieu's instability threshold was found and confirmed by computer simulation. The 'heating' in collisions with buffer gas molecules leads to ion decomposition. In the case of multicharged ions, fragments with an m/z value higher than that of the parent ion have a stable motion and can be analysed by an O-TOFMS. Fragmentation of bradykinin and beta-endorphin molecular ions having a selected charge state is demonstrated. The spectra show clear 'ladder' structure. The phenomenon may be used as an alternative to tandem mass spectrometry (MS/MS) for molecule structure analysis.