Concepedia

Publication | Open Access

Comparison of Three Different Pelvic Circumferential Compression Devices: A Biomechanical Cadaver Study

65

Citations

25

References

2010

Year

Abstract

The Pelvic Binder, SAM Sling, and T-POD provided sufficient reduction in partially stable and unstable (Tile type-B1 and C) pelvic fractures. No undesirable overreduction was noted. The pulling force that was needed to attain complete reduction of the fracture parts varied significantly among the three devices, with the T-POD requiring the lowest pulling force for fracture reduction.