IEEE Transactions on Biomedical Circuits and Systems · 2011 · 52 citations · 22 references
Life SciencesBiomedical EngineeringElectrical EngineeringNeural EngineeringProstheticsVestibular DisordersImplant TechnologyVestibular FunctionVestibular NeuroscienceVestibular PhysiologyVestibular SurgeryVestibular SystemData ProcessorVestibular CircuitsVestibular ImplantationVestibular Neurophysiology
This paper presents a multichannel stimulator ASIC for an implantable vestibular prosthesis. The system features versatile stimulation management which allows fine setting of the parameters for biphasic stimulation pulses. To address the problem of charge imbalance due to rounding errors, the digital processor can calculate and provide accurate charge correction. A technique to reduce the data rate to the stimulator is described. The stimulator ASIC was implemented in 0.6-μ m high-voltage CMOS technology occupying an area of 2.27 mm(2). The measured performance of the ASIC has been verified using vestibular electrodes in saline.
22
Thresholds for the perception of whole body angular movement about a vertical axis.
A. J. Benson, E. C. Hutt, S. F. Brown · PubMed · 1989 · 292 citations
Medicine, Neuroscience, Psychology +17
Charles C. Della Santina, Americo A. Migliaccio, Amit H. Patel · IEEE Transactions on Biomedical Engineering · 2007 · 172 citations · Full text