2005 IEEE Instrumentationand Measurement Technology Conference Proceedings · 2006 · 13 citations · 5 references
Natural SciencesBiomedical EngineeringPhysicsAcousticsBioinstrumentationComputational AcousticsAcoustic LocalizationAcoustic MeasurementAcoustic ModelingMedical InstrumentationSound LocalizationMicrophone Array LocalizationMicrophone Array MeasurementsMicrophone Array ModelingMicrophonesSound Source LocalizationSource LocalizationSource Localization Modeling
This paper compares two methods of source localization in the human thorax using a multisensor stethoscope. The multisensor stethoscope measurement technique consists of several stethoscope chest pieces simultaneously recording and storing audio data for computer-aided analysis. Localization method A finds the source location by measuring the relative power at different points in the field and identifying the location of maximum power. Localization method B finds the source by solving a system of equations based on the difference in time of arrival of signals at different microphones and the measured microphone positions. To illustrate this concept, a circular array of microphones equally spaced in one plane is assumed to be in a free-field homogeneous medium. Simulations, conducted in Matlab, as well as verification results are presented. Two in vitro verification models were constructed, and preliminary trials were done on a human test subject
5
Acoustic Imaging of the Human Chest
Martin Kompis, Hans Pasterkamp, George R. Wodicka · CHEST Journal · 2001 · 135 citations
Sound transmission in the lung as a function of lung volume
T. Bergstresser, D. Ofengeim, Andrey Vyshedskiy et al. · Journal of Applied Physiology · 2002 · 62 citations