Concepedia

Publication | Open Access

Improved Flagging Rates on the Sysmex XE-5000 Compared With the XE-2100 Reduce the Number of Manual Film Reviews and Increase Laboratory Productivity

47

Citations

16

References

2011

Year

Abstract

Hematology analyzers generate suspect flags in the presence of abnormal cells. False-positive rates for flags are high on all analyzers. Sysmex, Kobe, Japan, has developed new software for its XE-5000 with improved algorithms for flagging blast cells, abnormal lymphocytes or lymphoblasts, and atypical lymphocytes. This study evaluated the efficiency of these flags in 1,002 samples. The XE-5000 was compared with the XE-2100 (Sysmex) and microscopic examination of cell morphologic features. On the XE-2100, the blast flag demonstrated 90 false-positives, 13 true-positives, and 3 false-negatives. The values on the XE-5000 were 27 false-positives, 14 true-positives, and 2 false-negatives. The abnormal lymphocyte/lymphoblast flag was assessed with the atypical lymphocyte flag. The XE-2100 showed 114 false-positives, 23 true-positives, and 20 false-negatives, and on the XE-5000, there were 45 false-positives, 22 true-positives, and 21 false-negatives. This more specific flagging reduces the number of films that require manual review.

References

YearCitations

Page 1