Airflow efficacy of ballpoint pen tubes: a consideration for use in bystander cricothyrotomy

David M. Owens, B. Greenwood, A. Galley, A. Tomkinson, S L Woolley

Emergency Medicine Journal · 2010 · 15 citations · 19 references

Concepts

Abstract

OBJECTIVE: To examine the suitability of commonly available ballpoint pens as a substitute emergency tracheostomy tube. METHODS: Commonly available ballpoint pens were examined and compared against two standard cricothyroidotomy sets. The pens were evaluated for dimensions, speed of construction of a temporary tracheostomy tube and airway resistance with differing flow rates. RESULTS: Internal diameters of the pens varied considerably. Time taken to construct a temporary tube ranged from 3 to 170 s, and in the majority of pens the airway resistance increased dramatically as the airflow rate increased. CONCLUSION: Contrary to popular belief, the majority of ballpoint pens appear unsuitable for use as a substitute tracheostomy tube. In this study only two pens fulfilled the criteria for use: the Baron retractable ballpoint and the BIC soft feel Jumbo.

References

19