Journal of Neurology Neurosurgery & Psychiatry · 1980 · 238 citations · 4 references
EngineeringBioengineeringNeuroscienceElectrophysiologyMotor ControlMotor NeuroscienceElectromyogram AnalysisMuscle Signal ProcessingElectromyographyElectromyography InstrumentationElectromyography Signal AcquisitionElectromyography Signal InterpretationMotor Unit ActivationMuscle ContractionSkeletal Muscle Contraction
A new electromyographic technique for the study of the motor unit is described. A modified single fibre EMG electrode is used. The electrical activity obtained by the electrode shaft during voluntary muscle contraction is averaged after triggering from a single muscle fibre action potential and the contribution from one motor unit is thus extracted. This "Macro EMG" signal gives information about the whole motor unit in contrast to the regional or focal electrical activity measured by conventional or single fibre EMG recordings, respectively.
4
Erik Stålberg, Martin S. Schwartz, Barbara Thiele et al. · Journal of the Neurological Sciences · 1976 · 68 citations
Life Sciences, Anatomy, Physiology +11