IEEE Transactions on Biomedical Engineering · 2003 · 121 citations · 12 references
EngineeringElectrical EngineeringSurgeryAntennasMicrowave EngineeringRadiation ShieldingAntenna DesignAntenna RadiationAntenna SimulationMicrowave AntennasMinimally Invasive SurgerySmall AntennasCompact Antenna DesignEngineering SimulationsExperimental Results InterpretationImplantable Antenna DesignImplantable AntennasMicrowave Antenna DesignMiniaturized AntennasSystem Response Analysis
We present a new coaxial antenna for microwave interstitial coagulative therapy, working at 2450 MHz and endowed with a miniaturized sleeve choke in order to reduce back heating effects and make the system response less dependent on the antenna insertion depth into the tissue; the way the choke is implemented makes the overall transversal size minimum and allows small adjustments of the choke section length even during operation. We describe the main technical features of the antenna and show experimental results clearly proving the choke effectiveness. Numerical simulations well agree with experimental data, confirming the suitability of the proposed device for minimally invasive medical applications.
12
Field Computation by Moment Methods
D. S. Jones · The Computer Journal · 1969 · 3.5K citations · Full text
Robert D. Nevels, G. D. Arndt, G. Raffoul et al. · IEEE Transactions on Biomedical Engineering · 1998 · 104 citations
Engineering, Biomedical Engineering, Electrical Engineering +17