IEEE Transactions on Medical Imaging · 1991 · 100 citations · 18 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic Nuclear MedicineDiagnostic RadiologyBrain ImagingMedical ImagingNuclear Medicine ImagingPositron Emission TomographyRadiation Imaging SystemsScatter CorrectionAxial ResolutionCoincidence DetectionDeconvolutionNeurologic Pet ImagingPet Detector TechnologyPositron Emission Tomography DetectorsPositron Emission Tomography InstrumentationScatter Background Correction
Initial experience is reported with the Scanditronix PC 2048-15B, a 15-slice positron emission tomography (PET) system using multicrystal/multiphoto-multiplier modules to obtain high spatial resolution. Random and scattered events are reduced using an orbiting (68)Ge rod source for transmission scans by only accepting coincidence lines which intersect the instantaneous position of the source. Scatter correction of the emission data is removed with a deconvolution kernel, random and dead-time correction by the use of observed singles rates. The peak count rates are 11.7/20.0 Kcps for the direct cross slices at concentrations of 4.5/5.1 muCi/cc. respectively. Over the radial range 0-9 cm from the ring center, radial transverse resolution is 4.6-6.4 mm, aid tangential transverse resolution is 4.6-5.1 mm using a Hanning filter. Over the same range, axial resolution varies from 6.1-6.2 mm in direct slices and from 5.4-7.1 mm in cross slices. This near-isotropic resolution allows collection of image volume data with no preferred direction for signal averaging errors.
18
Brain blood flow measured with intravenous H2(15)O. II. Implementation and validation.
M. E. Raichle, W. R. Wayne Martin, Peter Herscovitch et al. · PubMed · 1983 · 1K citations