Optics Express · 2014 · 124 citations · 28 references
EngineeringElectrical EngineeringOpticsOptoelectronicsPhotonicsPolarization OpticsPolarization-based ImagingCircular PolarizationDevice FabricationElectrical Instrument CalibrationImaging PolarimetryInstrument CalibrationPolarimeter DesignPolarimetric Instrument DesignPolarimetry Instruments
In this paper, a full-Stokes imaging polarimeter operating at 580 nm using an array of elliptical polarizers is presented. The division-of-focal-plane polarimeter utilizes a set of four optimized measurements which represent a regular tetrahedron inscribed in the Poincaré sphere. Results from the device fabrication, instrument calibration and characterization are presented. The performance of the optimized full Stokes polarimeter, as defined by size of the standard deviation of the degree of circular polarization, is found to be approximately five times better than the performance of the simple full-Stokes polarimeter.
28
Review of passive imaging polarimetry for remote sensing applications
J. Scott Tyo, Dennis L. Goldstein, David B. Chenault et al. · Applied Optics · 2006 · 1.7K citations
Optimization of retardance for a complete Stokes polarimeter
Derek Sabatke, Michael R. Descour, Eustace L. Dereniak et al. · Optics Letters · 2000 · 354 citations · Full text