Sensors · 2008 · 24 citations · 30 references
Natural SciencesChemistryToxicologyAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectrochemical DetectionEnvironmental Analytical ChemistryFish ToxicologyHeavy Metal ContaminationInstrumental AnalysisDetection Limit AssessmentElectrochemical BehaviorElectrochemical MeasurementsGold ElectrodesHeavy Metal DetectionInstrument Detection LimitMercury Speciation AnalysisMethod Detection Limit
A simple method for manufacturing gold working electrodes for chronopotentiometric stripping measurements from recordable CD-R's is described. These gold electrodes are much cheaper than commercially available ones. The electrochemical behavior of such an electrode and the working parameters for mercury determination by chronopotentiometric stripping analysis were studied. Detection limit was 0.30 μg Hg/L and determination limit was 1.0 μg Hg/L for a deposition time of 600 s. Using the developed working electrodes it was possible to determine the total mercury in fish samples. A method for fish sample digestion was developed by using a mixture of fuming nitric acid and both concentrated sulfuric and hydrochloric acids. The recovery degree for a known amount of mercury introduced in the sample before digestion was 95.3% (n=4).
30
Gold Electrodes from Recordable CDs
Lúcio Angnes, Eduardo M. Richter, Márcio A. Augelli et al. · Analytical Chemistry · 2000 · 160 citations
Natural Sciences, Materials Engineering, Materials Science +12