Publication | Open Access
Reliability and Sensitivity of a Wrist Rig to Measure Motor Control and Spasticity in Poststroke Hemiplegia
34
Citations
35
References
2008
Year
The means of all tests differed between impaired and unimpaired participants except for range of movement into flexion, the FAI, and the FMI. Repeatability coefficients for each test are presented as benchmark values for use in future trials in which the wrist rig tests may be used to detect change. Test-retest reliability was excellent in the impaired group (ICC = 0.88-0.98) and poor to excellent in the unimpaired group (ICC = 0.06-0.89). The Bland-Altman ranges showed no bias between assessors, and that the interassessor variability was similar to that between repeats by the same assessor for most tests.