2006 IEEE Nuclear Science Symposium Conference Record · 2006 · 30 citations · 4 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDetector TechnologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationPositron Emission TomographyPositron Emission Tomography PhysicsInterferenceMagnetic Resonance Imaging AcquisitionPet Detector TechnologyPet-magnetic Resonance ImagingSystem Design Optimization
Features of a positron emission tomography-magnetic resonance (PET-MR) system in which a novel, superconducting magnet with a 'split' is combined with a multi-ring PET detector based on a microPETreg Focus 120 have been investigated. Those features include the impact on PET performance of: modification of the PET detector block, using a position-sensitive PMT in a nominal magnetic field of 10 Gauss (1mT), and - most significantly $operating the MR system whilst PET data is acquired. No effect on PET performance arising from MR radiofrequency (RF) or gradient pulses has been observed. RF interference between the PET electronics and MR signal has been monitored by acquiring MR data with the PET modules in situ and operational. That RF interference can be minimized by careful system design. The complete PET ring is now being constructed.
4
Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI.
Bernd J. Pichler, Martin S. Judenhofer, Ciprian Catana et al. · PubMed · 2006 · 397 citations
Simultaneous PET and MR imaging
Yiping Shao, Simon R. Cherry, Keyvan Farahani et al. · Physics in Medicine and Biology · 1997 · 373 citations