Evaluation of Strain at the Terminal Abutment Site of a Fixed Mandibular Implant Prosthesis During Cantilever Loading

48

Citations

11

References

1993

Year

Abstract

The results of this study indicate that an optimum biomechanical environment should exist when cantilever spans exceeding 7 mm are planned regardless of the number of supporting abutments. Strain transmitted to the crestal bone can be decreased by maximizing the number and anterior-posterior spread of supporting fixtures while minimizing the distance between the distal abutment and its adjacent abutment.