Concepedia

Publication | Open Access

Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter

80

Citations

16

References

2011

Year

Abstract

A characteristic of the software algorithm reduces the frequency of saturations of 87-90% and increases the frequency of higher values returned by the Masimo SET Radical pulse oximeter. This effect, which remains within the recommended standards for accuracy, is removed by installing revised software (board firmware V4.8 or higher). Because this observation is likely to influence oxygen targeting, it should be considered in the analysis of the oxygen trial results to maximise their generalisability.

References

YearCitations

Page 1