Publication | Open Access
The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
197
Citations
6
References
1997
Year
The mechanical motions used in this study significantly affected oximeter function, particularly when the sensors were connected during motion, which requires signal acquisition during motion. The error and dropout rate performance of the Masimo was superior to that of the other two instruments during all test conditions. Masimo uses a new paradigm for oximeter signal processing, which appears to represent a significant advance in low signal-to-noise performance.