Radiology · 2010 · 20 citations · 3 references
MedicineRadiologySurgeryDiagnostic ImagingDiagnostic RadiologyPeripheral Nerve SurgeryHand SurgeryThermal Injury PathophysiologyUpper Extremity Nerve SurgeryWrist DisordersWrist InjuriesAdverse Event DetectionCompartment Syndrome DetectionHand Compartment SyndromeHigh-frequency Electromagnetic FieldsMagnetic Resonance Imaging ContraindicationsNeed For Surgical InterventionThermal Tissue EffectsWrist Imaging
Magnetic resonance (MR) imaging is generally considered a safe procedure. Contraindications include the presence of foreign objects in or on the body, which may be subject to electromagnetic fields associated with the MR system. Most of these objects are well known and are routinely screened for prior to the procedure. The authors report an unusual adverse event that appears to have been caused by a unique combination of factors involving an identification bracelet, an item not previously known to present any risks. To the authors' knowledge, this is the first report in the literature of a severe electrical thermal burn that required surgical intervention. Identification bracelets may need to be removed or padded to prevent direct contact with the patient's skin during all MR imaging examinations for patients unable to communicate, such as those requiring sedation or general anesthesia. (c) RSNA, 2010.
3
Michael V. Knopp, Marco Essig, Jürgen Debus et al. · Radiology · 1996 · 59 citations
MRI induced fourth-degree burn in an extremity, leading to amputation
Josef Haik, Daniel Simon, Ariel Tessone et al. · Burns · 2008 · 28 citations