Determination of the volumetric mass transfer coefficient (<i>k</i><sub><i>L</i></sub><i>a</i>) using the dynamic “gas out–gas in” method: Analysis of errors caused by dissolved oxygen probes

Lauren Tribe, Cédric Briens, Argyrios Margaritis

Biotechnology and Bioengineering · 1995 · 132 citations · 19 references

DOIFull text

Open access

Concepts

Abstract

There are many dynamic methods for measuring the volumetric mass transfer coefficient. The "gas out-gas in" method can directly determine the volumetric mass transfer coefficient in a bioreactor system and provide estimates of the volumetric microbial oxygen uptake rate and the average oxygen saturation concentration at the gas-liquid interface. The errors on these parameters are large if the dissolved oxygen probe response time is not considered. For reliable measurements, deconvolution of the oxygen probe measurements must be made. (c) 1995 John Wiley & Sons, Inc.

References