SMART: The microstent's ability to limit restenosis trial

Richard R. Heuser, Alejandro Lopez, Richard Kuntz, Lawrence A. Reduto, Rodney S. Badger, Patrick L. Coleman, Patrick L. Whitlow, L.A. Iannone, Robert D. Safian, Alan C. Yeung,

Catheterization and Cardiovascular Interventions · 2001 · 16 citations · 15 references

Abstract

In this randomized, prospective, multicenter trial (n = 661) of patients with de novo or restenotic coronary lesions, 330 patients received the MicroStent(R) II (MSII), and 331 received the Palmaz-Schatz (PS) stent. The short-term procedural success rates were 94.4% and 95.7%, respectively (P = 0.47). The 30-day cumulative incidence of major adverse events [death, myocardial infarction, CVA, target lesion revascularization (TLR)] was 6.4% for the MSII and 4.5% for the PS stent (P = 0.31). The in-stent binary restenosis rate at 6 months was 25.2% for the MSII and 22.1% for the PS stent (P = 0.636). Using Kaplan-Meier estimates, the incidence of clinically driven TLR was 8.9% for the MSII and 9.2% for the PS stent at 180 days; at 270 days, it was 12.8% and 12.1%, respectively (P = 0.83). MSII and the PS stents were comparable with respect to short-term procedural success, complications, and late clinical and angiographic restenosis.

References

15