Publication | Open Access
Impact of Freezing Time and Balloon Size on the Thermodynamics and Isolation Efficacy During Pulmonary Vein Isolation Using the Second Generation Cryoballoon
49
Citations
13
References
2015
Year
The efficacy of 3-minute ablation was not significantly different from 4-minute ablation in dogs. The 23-mm cryoballoon had a greater cooling effect than the 28-mm cryoballoon for small PVs, but showed narrower ablative lesions in the left atrial antrum.