Journal of Environmental Monitoring · 1999 · 25 citations · 10 references
Natural SciencesToxicologyAir PollutionAnalytical ChemistryAtmospheric ChemistryAir Pollution ChemistryAir Pollution MeasurementAir SamplingContaminant AnalysisEnvironmental AnalysisEnvironmental Analytical ChemistryFilter-based Air SamplingFormaldehydeHigh-performance Liquid ChromatographyHydrazone FormationMethod Detection LimitSampling RatesUltraviolet-visible DetectionUv-visible Quantitative Analysis
A new method utilizing the diffusive sampling of formaldehyde in air has been developed. Formaldehyde is sampled with the use of a glass fiber filter impregnated with N-methyl-4-hydrazino-7-nitrobenzofurazan (MNBDH) and phosphoric acid. The formaldehyde hydrazone formed is desorbed from the filter with acetonitrile and determined by high-performance liquid chromatography (HPLC) with UV/visible detection at 474 nm. The sampling rate was determined to be 24.7 mL min-1 with a relative standard deviation of 7% for 48 experiments. The measured sampling rates were not dependent on the formaldehyde concentration (0.1-1.0 mg m-3), sampling time (15-482 min) or relative humidity (20-85%). The detection limit was 70 micrograms m-3 for a 15 min sampling period and 2 micrograms m-3 for an 8 h sampling period.
10
Modified pararosaniline method for the determination of formaldehyde in air
R.R. Miksch, D. W. Anthon, L.Z. Fanning et al. · Analytical Chemistry · 1981 · 157 citations
Jan Olof. Levin, Kurt Andersson, Roger Lindahl et al. · Analytical Chemistry · 1985 · 119 citations
David F. Smith, Tadeusz E. Kleindienst, Edward Hudgens · Journal of Chromatography A · 1989 · 89 citations