Concepedia

Publication | Closed Access

Initial Experience with 1.5‐mm<sup>2</sup> High Impedance, Steroid‐luting Pacing Electrodes

22

Citations

15

References

1996

Year

Abstract

The present study validates that smaller surface (i.e., 1.5 mm2) steroid-eluting electrode designs offer excellent pacing and sensing performance with significantly higher pacing impedances. Although questions remain as to the cause of late exit block in two patients in this series, this relatively small surface electrode design offers promise toward achieving greater pacing efficiency and a theoretical 13%-16% (minimum) enhancement in permanent pacemaker longevity.

References

YearCitations