Boneloc® cemented total hip prostheses: Loosening in 28/43 cases after 3-38 months

Per Riegels-Nielsen, Lilli Sørensen, H. M. Andersen, S Lindequist

Acta Orthopaedica Scandinavica · 1995 · 29 citations · 8 references

DOIFull text

Open access

Concepts

Abstract

We report our early results with 43 total hip prostheses (25 Charnley and 18 LMT) inserted during 1991 with Boneloc cement. The indication was primary arthrosis (38) and rheumatoid arthritis (5). After 18 (3-38) months, 28 stems were loose and 18 hips have been revised 1-4 years after primary surgery.

References

8