Comparison of ion chamber and TLD dosimetry in mammography
16
Citations
0
References
1981
Year
EngineeringMeasurementBiomedical EngineeringTreatment VerificationCalibrationBreast ImagingRadiation Therapy PlanningExposure Vs DepthMemorial Mammography ChamberInstrumentationRadiation ImagingRadiation OncologyIonization Chamber MethodNuclear MedicineRadiologyHealth SciencesMedical ImagingTld DosimetryUltrasoundRadiographic ImagingDosimetryBreast Cancer
An ionization chamber method has been developed to measure exposure vs depth in a uniform BR 12 "average breast" phantom. It employs a Memorial mammography chamber for exit exposure measurements; resulting data is then corrected for backscatter as well as for the exceptionally thin window of this chamber. A careful comparison has then been made with relative exposure vs depth curves obtained using TLD at several mammography beam qualities, for identical exposure factors and SSD values. Use of a correction for residual and background TL signals significantly improved agreement between TLD and ion chamber curves in the 28 to 35 kVp/0.03 mm Mo range of beam quality. Agreement was within +/- 5% for the Mo target tube, but TLD readings were 4%--8% higher than ion chamber values for the W/Mo target tube. At Xeromammography energies (45 kVp/1.6 mm Al), corrected TLD curve readings were 6% higher at depth than ion chamber curve values. TLD meaurements with 28 to 35 kVp/0.03 mm Mo beams tend to underestimate dosage to the midbreast parenchyma. For example, in a 5 cm "average breast", the underestimation ranges from 2%--10% for corrected, 10%--16% for uncorrected TLD readings.