Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays

Isabelle Strubi-Vuillaume, Valentine Carlier, Catherine Obeuf, Francis Vasseur, J-Claude Maury, Patrice Maboudou, Jacques Mangalaboyi, Alain Durocher, David Launay, Christian Noël,

Annals of Clinical Biochemistry International Journal of Laboratory Medicine · 2015 · 11 citations · 2 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: Use of a hospital pneumatic tube system may be associated with measurement errors. METHODS: A venous blood sample was collected from 79 patients into a pair of lithium heparin tubes; one tube was sent to the laboratory by porter and the other was sent via the pneumatic tube system. Plasma lactate dehydrogenase concentrations were then assayed. RESULTS: Lactate dehydrogenase concentrations were overestimated (median bias: 18.8%) when evacuated vacuum lithium heparin tubes were sent by pneumatic tube system. This bias was reduced by bubble-wrapping the standard lithium heparin tube or using Monovette lithium heparin tubes in aspiration mode (median bias: +8.7% and -0.3%, respectively). CONCLUSIONS: Cushioning and aspiration-mode sampling may limit pneumatic tube system-associated overestimation of lactate dehydrogenase concentrations.