Publication | Open Access
Performance Evaluation of Two Published Closed-loop Control Systems Using Bispectral Index Monitoring
129
Citations
0
References
2004
Year
When simulating closed-loop control of BIS using propofol, the use of a patient-individualized, model-based adaptive closed-loop system with effect site control resulted in better control of BIS compared with a standard proportional integral derivative controller with plasma site control. Even under extreme conditions, the modeled-based controller exhibited no behavioral problems.