Intravascular ultrasound tissue harmonic imaging in vivo
11
Citations
16
References
2005
Year
Unknown Venue
Medical UltrasoundEngineeringHealth SciencesMedical ImagingPower UltrasoundVascular ImageIvus CatheterBiomedical ImagingTissue Harmonic ImagingElastographyElectrophysiologyBiomedical EngineeringUltrasoundConventional Ivus CatheterBlood Flow MeasurementRadiologyCardiovascular Imaging
The feasibility of tissue harmonic imaging with a prototype intravascular ultrasound (IVUS) system using a conventional single-element rotating IVUS catheter was investigated. Hydrophone measurements showed fundamental peak pressures of 1.9 MPa on-axis of such a catheter in water. The second harmonic signal at 40 MHz built up to 0.55 MPa, corresponding to 11 dB below the peak fundamental pressure. Experiments performed in a tissue-mimicking phantom and in a rabbit aorta in vivo showed the feasibility of tissue harmonic imaging with a conventional IVUS catheter. When we applied averaging of four neighboring RF-lines in the in vivo acquisition, the signal-to-noise ratio of the harmonic image increased to 25 dB.