IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 2002 · 17 citations · 18 references
MedicineBiomedical EngineeringCardiologyBiomedical ImagingDiagnostic ImagingUltrasoundCardiac ImagingCardiac UltrasonographyCardiac UltrasoundCardiovascular ImagingMedical ImagingMedical Imaging InstrumentationFractional BandwidthInterventricular SeptumReal-time ImagingSpatial Resolution MeasurementTransducer Array DesignTransducer CharacterizationTransducer DesignVolumetric Imaging
The design, fabrication, and characterization of a 7 MHz, two-dimensional (2-D) array transducer built on a silicon substrate is described. The array fits inside a 9-French (2.9 mm O.D.) catheter for use in real-time intracardiac volumetric imaging. The -6 dB fractional bandwidth of the transducer is 30%, the 50 ohm pitch-catch insertion loss is 78 dB, and the interelement crosstalk is -25 dB. Real-time volumetric images in phantoms and in-vitro images of a sheep heart have been acquired yielding measured spatial resolution of 2 mm at a depth of 1 cm. The cardiac structures imaged include ventricular chambers, interventricular septum, mitral and tricuspid valves and real-time 3-D rendered volumes of the tricuspid valve in the open and closed position.
18
Radiofrequency catheter ablation guided by intracardiac echocardiography.
Edward Chu, Adam Fitzpatrick, M C Chin et al. · Circulation · 1994 · 143 citations · Full text