Applied Spectroscopy · 1984 · 34 citations · 40 references
Natural SciencesChemistryPhysicsAnalytical ChemistryMass Spectrometric AnalysisPlasma PhysicsAerosol ScienceNebulizer SystemsAerosol InstrumentationAnalytical Detection LimitsGas FlowInductively Coupled Plasma SpectrometryJet NebulizationJet NebulizersNebulizer Device DesignSample Introduction Systems
A novel, simple device for the production of liquid aerosol has been constructed and evaluated. In operation of this device, a sample solution is forced under pneumatic or mechanical pressure through an orifice of 25–60 micrometers in diameter. The resulting high-velocity stream of solution is directed onto a solid surface, thereby generating a fine aerosol. Details of nebulizer construction and operation are discussed and factors affecting nebulizer performance are examined. The utility of this nebulizer as a sample-introduction device for inductively coupled plasma spectrometry is explored. When coupled with an ICP, the jet-impact nebulizer compares favorably with a conventional pneumatic nebulizer in terms of precision, linearity, detection limits, and efficiency. Moreover, aerosol production is independent of any gas flow, making the device potentially useful in low-gas-flow plasmas. However, clogging of the nebulizer orifice with particulate matter can be a problem unless solutions are pre-filtered.
40
K. W. Olson, W.J. De Haas, V. A. Fassel · Analytical Chemistry · 1977 · 206 citations
A. M. Gunn, David Millard, G. F. Kirkbright · The Analyst · 1978 · 149 citations