IEEE Journal of Solid-State Circuits · 1998 · 24 citations · 9 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsBiomedical ImagingDiagnostic ImagingBiomedical InstrumentationCmos Circuit DesignDetector TechnologyLow-noise Circuit DesignMedical ImagingMedical Imaging InstrumentationCharge DetectionCurrentsDetector Readout ElectronicsLow-noise AmplificationLow-noise DesignLow-noise Detector DesignNoise Performance AnalysisSolid-state Detectors
A multipurpose digital detector readout for medical imaging applications is presented. The readout is capable of measuring both current and charge, allowing a single detector array to perform imaging functions previously accomplished with two separate machines. The circuit employs a variable rate /spl Sigma//spl Delta/ analog-to-digital converter (ADC) to measure current over a 130-dB dynamic range in a 1 kHz band and resolve charge pulses down to 360 e/sup -/ at 100 000 events/s. Detector currents of up to 7 /spl mu/A and charge pulses as large as 25 fC can be measured. A low-noise charge sensing amplifier (CSA) is combined with digital pulse shaping to optimize the noise performance and flexibility of the charge measurements. Fabricated in an 1.2 /spl mu/m complimentary metal-oxide-semiconductor (CMOS), the circuit occupies 1.5 mm/sup 2/ and dissipates 11 mW/channel from a 5 V supply.
9
Description of a prototype emission-transmission computed tomography imaging system.
Thomas Lang, B.H. Hasegawa, Soo Chin Liew et al. · PubMed · 1992 · 175 citations