Non-contact SensingElectrical EngineeringMedical ElectronicsEngineeringElectrophysiological EvaluationBioelectronicsCapacitive Biopotential MeasurementsElectrophysiologyBiomedical EngineeringInstrumentationBioinstrumentationContact ElectrificationBiophysicsElectrochemistryElectrical InsulationStrong Motion Artifacts
Capacitive biopotential measurements suffer from strong motion artifacts, which may result in long time periods during which a reliable measurement is not possible. This study examines contact electrification and triboelectricity as possible reasons for these artifacts and discusses local triboelectric effects on the electrode-body interface as well as global electrostatic effects as common-mode interferences. It will be shown that most probably the triboelectric effects on the electrode-body interface are the main reason for artifacts, and a reduction of artifacts can only be achieved with a proper design of the electrode-body interface. For a deeper understanding of the observed effects, a mathematical model for triboelectric effects in highly isolated capacitive biopotential measurements is presented and verified with experiments. Based on these analyses of the triboelectric effects on the electrode-body interface, different electrode designs are developed and analyzed in order to minimize artifacts due to triboelectricity on the electrode-body interface.