Concepedia

Publication | Closed Access

Radiation Dose of Interventional Radiology System Using a Flat-Panel Detector

90

Citations

19

References

2009

Year

Abstract

We found that the average entrance doses of cineangiography and fluoroscopy in FPD systems were not significantly different from those in II systems. Hence, FPDs did not inherently reduce the radiation dose, although FPDs possess good detective quantum efficiency. Therefore, to reduce the radiation dose of cardiac interventional radiology systems, even FPD systems, practical measures are necessary.

References

YearCitations

Page 1