Influence of attachment wear on retention of mandibular overdenture
129
Citations
51
References
2006
Year
one-way anova, Scheffe post hoc and paired-samples t-tests (P < 0.05). Initially, studs had higher retention (4-11 N) than magnets (4.5-6 N) in axial direction. After the wear simulation, it had decreased from 76% to 48% for some of the studs and had become similar to the retention of magnetic attachments. Magnets had lower retention range (0.2-0.3 mm) than studs (0.5-1.1 mm). Studs provided similar or higher retention in paraxial directions than magnetic attachments both before and after wear simulation. Retentive properties of magnets decreased mostly with posterior rotational dislodgement. Retentive properties of stud overdenture attachments were less constant.