A head and neck hyperthermia applicator: Theoretical antenna array design

64

Citations

14

References

2007

Year

Abstract

A single optimum array setup suitable for all patients is difficult to define. Based on the results and practical limitations a setup consisting of two rings of six antennas with a radius of 20 cm and 6 cm array spacing is considered a good choice providing the ability to heat the majority of patients.