Concepedia

Publication | Open Access

Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study

41

Citations

25

References

2013

Year

Abstract

There is poor agreement between RBT and GAT in children with glaucoma. RBT frequently and significantly overestimates IOP. However, 'normal' RBT readings are likely to be accurate and may spare children an examination under anaesthesia (EUA). High RBT readings should prompt the practitioner to use another standard method of IOP measurement if possible, or consider the RBT measurement in the context of clinical findings before referring the child to a specialist clinic or considering EUA.

References

YearCitations