Journal of High Resolution Chromatography · 1985 · 33 citations · 5 references
EngineeringSupercritical Fluid ChromatographyChemistryChemical EngineeringGas ChromatographyAnalytical ChemistryLiquid ChromatographyMass Spectrometric DetectionChromatographySource TemperatureCo 2Supercritical FlowChromatographic AnalysisSimultaneous Flame IonizationIon SourceSupercritical Co2Mass SpectrometryMedicineDrug Analysis
Abstract A simple interface between a capillary supercritical fluid chromatograph and an Extranuclear Simulscan mass spectrometer is described. The SFC column is directly inserted into the ion source through the existing GC‐interface. The system is equipped with a splitting device which allows simultaneous EI/MS and flame ionization detection when CO 2 is used as the supercritical phase. The effect of source temperature and pressure on CO 2 clustering was studied for optimization of source conditions. The performance of the system was evaluated with a series of model compounds and standard mixtures.
5
T. L. Chester · Journal of Chromatography A · 1984 · 104 citations
Supercritical Fluid Chromatography, Supercritical Co2, Detection Artifacts +11