IEEE Transactions on Instrumentation and Measurement · 1990 · 58 citations · 7 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMetrologyCalibrationEquipment CalibrationMeasurement StandardsMicrowave MeasurementsDeconvolutionDynamic CalibrationInstrument CalibrationInstrument Channel CalibrationOscilloscope CalibrationOscilloscope-based MeasurementRecording Equipment CalibrationWaveform Digitization
The current methods used to calibrate oscilloscopes are briefly discussed, and the problems associated with oscilloscope vertical channel bandwidth testing are illustrated. A method is presented for the complete and correct calibration of an oscilloscope voltage channel. A solution that involves measuring and digitizing pulse signals and an NIST-developed deconvolution algorithm is shown. An example of the calibration of the vertical channel of a 20-GHz sampling oscilloscope is presented.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
7
Solutions of Ill-Posed Problems.
John B. Bell, V. Y. Arsenin · Mathematics of Computation · 1978 · 8.2K citations