Straw Detector for High Rate, High Resolution Neutron Imaging

Athanasios Athanasiades, N.N. Shehad, Christopher S. Martin, L. T. Sun, Jeffrey L. Lacy

2006 · 36 citations · 3 references

Concepts

Abstract

Instrument requirements in soon to be commissioned and planned neutron scattering facilities call for large area detectors that can sustain rates up to 10/sup 8/ cps and millimeter-level spatial resolution. Although /sup 3/He pressurized area detectors can provide good spatial resolution, sensitivity and gamma ray discrimination, this technology cannot achieve the required rates without further development. Moreover, achieving large detection areas with pressurized /sup 3/He technology is expensive because of the complexity of the pressure containing vessels required. We propose a detector technology, based on thin-walled straws, lined with a 1 /spl mu/m thick sputter coating of enriched boron carbide (/sup 10/B/sub 4/C). Neutrons converted in /sup 10/B generate charged particles that subsequently ionize the gas contained within each straw. Because the /sup 10/B/sub 4/C coating is very thin, efficient escape of the reaction products can be achieved. A panel detector consisting of several thousand close-packed individual straws, which are read out independently, can easily support high event rates. We present performance testing of two 50-straw prototypes (non-enriched B/sub 4/C), including detection efficiency, spatial resolution, and two-dimensional imaging. Each straw has a diameter of 4 mm and a length of 1 m. Additional tests of a single 50 cm long straw, 2 mm in diameter are also presented.

References

3