Publication | Open Access
Is there a need for routine testing of ICD defibrillation capacity?. Results from more than 1000 studies
21
Citations
10
References
2000
Year
Our experience demonstrates that postoperative tests of the defibrillation function of implantable cardioverter-defibrillators rarely reveal dysfunctions. As testing is unpleasant for the patient and not free of complications, tests might be restricted to those patients in whom a dysfunction is suspected and to those patients in whom class I or class III antiarrhythmic drugs have been added to the antiarrhythmic drug regimen.