Publication | Open Access
New disposable tubes for rapid and precise biomass assessment for suspension cultures of mammalian cells
48
Citations
8
References
2006
Year
The study validates a disposable microcentrifuge tube method for rapid biomass assessment in widely used mammalian cell lines. The tube features an upper loading chamber and a lower 5 µL capillary that collects cells during centrifugation, enabling packed cell volume to be expressed as a percentage of the sample volume. Across multiple examples, the packed cell volume method proved more precise, faster, and more reproducible than manual cell counting. © 2006 Wiley Periodicals, Inc.
We present a new approach for biomass assessment in cell culture using a disposable microcentrifuge tube. The specially designed tube is fitted with an upper chamber for sample loading and a lower 5 µL capillary for cell collection during centrifugation. The resulting packed cell volume (PCV) can be quantitatively expressed as the percentage of the total volume of the sample. The present study focused on the validation of the method with mammalian cell lines that are widely used in bioprocessing. Using several examples, the PCV method was shown to be more precise, rapid, and reproducible than manual cell counting. © 2006 Wiley Periodicals, Inc.
| Year | Citations | |
|---|---|---|