Publication | Open Access
Animal model to compare the effects of suture technique on cross-sectional compliance on end-to-side anastomoses
28
Citations
8
References
2001
Year
This model, even with some limitations, can be a reliable source of information improving the outcome of vascular anastomoses. The study demonstrates that suture technique has a substantial effect on cross-sectional anastomotic compliance of end-to-side anastomoses. Interrupted suture may maximise the anastomotic lumen and provides a considerably higher CSAC than continuous suture, that reduces flow turbulence, shear stress and intimal hyperplasia. The Heartflo anastomosis device is a reliable instrument that facilitates performance of interrupted suture anastomoses.